ClearanceBrief
Recall Z-2878-2020 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Ortho Clinical Diagnostics

VITROS XT3400 Chemistry System, Product code 6844458

DateJul 20, 2020
StatusTerminated
ReasonWhen processing MicroSlides, slides that utilize Slot 18 in the CM rotor may be mis-positioned. As a result, certain VITROS MicroSlides may produce erroneous results.
Root causeProcess control
ActionThe firm notified customers through Urgent Product Correction Notification letters on July 20, 2020. Customers were informed of the potential issue with incubator slot 18 of affected devices. Customers were provided with a table of the eight assays with biases which do not align with the assay's intended clinical use. Customers who had already been identified as experiencing this issue through their e-connected systems, or who did not have e-connected systems, received a letter stating that the recalling firm would be replacing the CM Rotor to resolve the issue. Customers who had e-connected
Quantitypending
DistributionWorldwide distribution - US Nationwide distribution and the countries of Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Poland, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K182063VITROS Chemistry Products CRBM Slides, VITROS Chemistry Products CREA Slides, VITROS Chemistry Products TBIL Slides, VIT
Ortho-Clinical Diagnostics, Inc., Oct 30, 2018