Recall Z-2885-2020 ยท OWB Interventional Fluoroscopic X-Ray System

Canon Medical System, USA, INC.

System Model: Ultimax-i, DREX-UI80 (DREX-UI80); Unit Model: HDR-08A Imager Processor; UDI (01)04987670100147

DateJul 8, 2020
StatusOpen, Classified
ReasonDuring a procedure, when images were acquired, and these images were transferred to PACS /image archival system, the number of images displayed were not the same as originally acquired.
Root causeSoftware design
ActionThe recall letters were sent via facsimile between 07/08-10/2020. The letter contains a return reply form that is to be completed by the customer and faxed to CMSU for retention. After approximately one month contact will be made with each non-responding location and a copy of the letter will be faxed to that location and the fax receipt will serve as proof of notification.
Quantity11 units
DistributionUS - CA, CT, ID, LA, MT, NV, OH, SC, WI and WV
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K181834Ultimax-i, DREX-UI80 (V1.60)
Canon Medical Systems Corporation, Jul 19, 2018