Recall Z-2715-2020 ยท LHN System, Radiation Therapy, Charged-Particle, Medical

Ion Beam Applications S.A.

Proteus 235 The Proton Therapy System - Proteus 235 (brand names: Proteus Plus and Proteus ONE) is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.

DateJul 17, 2020
StatusTerminated
ReasonIBA became aware of an issue to resume aborted irradiations because the plan to complete the treatment cannot be loaded.
Root causeSoftware design
ActionAn Urgent Medical Device Correction notification will be communicated to all impacted IBA customers (users of the device), starting July 23, 2020. This letter instructs the customer: If this situation were to occur, it is recommended to complete the aborted treatment in Standalone Mode (please refer to the steps described in annex I and to the part Using adaPTdeliver in Standalone Mode of the Proton Therapy System Clinical Users Guide). IBA will modify the software to remove conditions that make impossible to load the plan to complete the aborted treatment. The solution will be deployed on you
Quantity2 Units
DistributionUS: VA OUS: Belgium
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K163500IBA Proton Therapy System- Proteus 235 (Proteus One, Proteus Plus, Proteus TK2, Proteus Class)
Ion Beam Applications S.A., Aug 17, 2017