Recall Z-2874-2020 ยท CIG Diazo Colorimetry, Bilirubin

Ortho-Clinical Diagnostics

VITROS Chemistry Products BuBc Slides - in vitro diagnostic use only. VITROS Chemistry Products BuBc Slides quantitatively measure unconjugated bilirubin (Bu) and conjugated bilirubin (Bc) concentrations in serum and plasma, Product code: 8383051 (60 slide); 1612365 (18 slide)

DateJul 13, 2020
StatusTerminated
ReasonEtrombopag Interference for VITROS Chemistry Products: TBIL Slides, VITROS BuBc Slides and VITROS XT TBIL-ALKP Slides
Root causeOther
ActionOrtho Clinical Diagnostics issued URGENT PRODUCT CORRECTION NOTIFICATION on 13 July 2020, customer letter (Ref. CL2018-181) via FedEx overnight courier and/ or ORTHO PLUS e-Communications and/ or US Postal Service Priority Mail (for PO Boxes only) to all customers who received one of the affected products in the previous 18 months. Foreign affiliates were informed by email on 13 July 2020 of the issue and instructed to notify their consignees of the issue and their required action until the updated the IFU is available. Letter states reason for recall, health risk and action to take: Instr
Quantity80351:US=35113; OUS: OUS=1593941.1612365: US=20068; OUS= 35534
DistributionWorldwide distribution - US Nationwide distribution and the countries of Australia, Belgium, Bermuda HM08, Brazil, Canada L3R 4G5, China, Denmark, France, Germany, Italy, Mexico, Norway, Poland, Portugal, Russia, Singapore 609917, Spain, Sweden, The Netherlands, United Kingdom.
FDA recordOpen on fda.gov