Recall Z-2753-2020 ยท CHI Lipase-Esterase, Enzymatic, Photometric, Lipase

RANDOX LABORATORIES, LTD.

Randox Lipase Assay: Lipase Colorimetric R2 Substrate - Product Usage: A Lipase test system is a device intended for the quantitative in vitro determination of Lipase in human serum and plasma.

DateAug 4, 2020 (initiated Jun 26, 2020)
StatusTerminated
ReasonRandox have confirmed imprecision of quality control and patient samples when using specific Randox Lipase R2 Substrate batches.
Root causeUnknown/Undetermined by firm
ActionRandox sent an urgent medical device correction notification dated June 22, 2020 explaining the current failure with select lots of Lipase Colorimetric products. The firm advised customers to scrap all effected product, keep proof of destruction, review replacements needs, review results of patients generated using affected batches and return the firm's response card.
DistributionWorldwide distribution - US Nationwide distribution including in the states of Puerto Rico and WV.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.