ClearanceBrief
Recall Z-2786-2020 ยท HRS Plate, Fixation, Bone

Flower Orthopedics Corporation

Calcaneus Plate, Extensile, Medium, Right, Catalog Number CPF 202 - Product Usage: intended to be used for internal fixation of fractures and reconstruction of bones of the rearfoot, including the calcaneus.

DateAug 12, 2020 (initiated Jul 8, 2020)
StatusTerminated
ReasonThe contract manufacturer of the calcaneus plate inadvertently used the incorrect material to manufacture the plates (a titanium alloy rather than pure titanium).
Root causeMixed-up of materials/components
ActionThe Regional Business Director notified the customer by telephone and informed them that there was an issue with the newly released calcaneus plates. The 6 units were requested to be returned to Flower Orthopedics.
Quantity25
DistributionUS Nationwide distribution including in the states of AZ, FL, and OH.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133930FLOWER REAR FOOT PLATING SET, FLOWER CALCANEUS PLANTING SET, FLOWER REARFOOT PLATING SET
Flower Orthopedics Corporation, Mar 27, 2014