ClearanceBrief
Recall Z-2894-2020 ยท CAL Laryngoscope, Non-Rigid

Karl Storz Endoscopy

Flexible DCI Intubation Fiberscope, Part: 11301BND1, with Instruction Manual: Z18445US - Product Usage: are intended to be used for examination and visualization of a patient s upper airway and for aiding placement of an endotracheal tube.

DateJul 27, 2020
StatusOpen, Classified
ReasonUpdated sterilization instructions are being provided to ensure 1) a specific sterilization tray is not used, and 2) ethylene oxide sterilization is never used on certain reusable flexible intubation endoscopes with total lumen lengths greater than 850mm. Incompletely sterilized endoscopes might introduce infectious agents into patient airways, resulting in infection.
Root causeDevice Design
ActionOn 07/27/2020, Urgent Medical Device Correction notices were mailed to customers. Customers were informed that this labeling change corrects the instructions for sterilization of the impacted endoscopes using STERRAD or EtO modalities. Do not use any impacted endoscope until you have confirmed that it has been appropriately sterilized in accordance with the instructions in the correction notice or via another method specified in instructions for use with updated sterilization information. When sterilizing one of the impacted flexible intubation endoscopes with the STERRAD NX Advanced Cycle or
Quantity463
DistributionWorldwide distribution - U.S. Nationwide distribution including in the states of CA, NY, PA, AZ, OH, NJ, FL, NC, IL, SC, MA, IA, ND, WI, UT, AR, GA, WA, KY, CO, TX, TN, AL, LA, NE, WV, MO, MN, VA, MT, CT, MI, IN, NM, NV, ME, VT, NH, WY, DC, DE, OR, OK, MD, ID, AK, MS, KS, HI, RI, SD and the country
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.