Recall Z-2769-2020 ยท FCG Biopsy Needle

Covidien Llc

Beacon EUS Access System Preloaded Access 135degree Needle sterile, single-use, endoscopic ultrasound device consisting of a sharp stylet and an 18.5-gauge, echo-enhanced, tip-shaped, nitinol cannula intended to pierce tissue and then act as an access channel to facilitate guidewire placement to locations of the gastrointestinal tract.

DateAug 5, 2020 (initiated Jun 26, 2020)
StatusOpen, Classified
ReasonThis voluntary recall is being conducted following three customer reports of the device cannula detaching during a procedure. In one case, the detached cannula was not retrieved during the procedure. Potential risks associated with a detached cannula include delay in treatment, biliary leak, pain and retained device fragments.
Root causeUnder Investigation by firm
ActionMedtronic sent a letter notification to customers dated July 06, 2020 recalling effected lots of product. This letter instructed customers to discontinue use of the product and quarantine effected inventory. Customers were also instructed to return effected product and complete recall response forms accordingly.
Quantity269 units
Distributionworld wide distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K180037Beacon EUS Access System
Covidien, LLC, Apr 26, 2018