Recall Z-2884-2020 ยท PDQ Neurosurgical Nerve Locator

The Magstim Company Limited

Neurosign V4 Intraoperative Nerve monitor family of devices.

DateJun 23, 2020
StatusTerminated
ReasonReports that manipulation of the cable which connects the pre-amplifier to the nerve monitor, and the cable which connects the pre-amplifier to the stimulator pod, may cause damage to the internal construction of the cables.
Root causeDevice Design
ActionOn June 23 2020, the firm sent a urgent field safety notice to its costumers. What you need to do: - Identify all affected products at your facility - Raise awareness to all users of the device(s) within your organisation - If you have forwarded any affected product to other organisations, please forward them a copy of this notice, and advise us accordingly on the acknowledgement form - Complete and return the attached acknowledgement form, so we can contact you to arrange replacement of the cable. - Users should continue to routinely inspect all devices for damage or signs of wear before use,
Quantity49 devices
DistributionDomestic: MN International: UK, Republic of Ireland, Spain, Netherlands, Canada. (Technomed Europe distributed to: Brazil, Spain, Belgium, Chile, Greece, Poland, New Zealand, Pakistan, Portugal and Thailand.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K181559Neurosign V4 Intraoperative Nerve Monitor
Magstim Company, Ltd., Nov 9, 2018