Recall Z-2824-2020 ยท FRN Pump, Infusion
CareFusion 303, Inc.
Alaris PCA Module Model 8120 - Product Usage: is intended to provide trained healthcare caregivers a way to automate the programming of infusion parameters, thereby decreasing the amount of manual steps necessary to enter infusion data.
| Date | Aug 19, 2020 (initiated Jun 30, 2020) |
| Status | Open, Classified |
| Reason | LED display, which provides infusion or patient monitoring values, on the module may have segments that appear dim. |
| Root cause | Device Design |
| Action | On 06/30/2020, Urgent Medical Device Recall Notifications will be mailed to Directors of Biomedical Engineering, Nursing, Risk Management, Environmental Services. Actions for Clinical User: 1. If the end user observes a dim segment on the module, replace the module when feasible. There is no impact to the infusion. 2. Complete and return the Customer Response Card. Actions by Biomedical Engineering: If the module has a dim segment, please contactrecalling firm at 1-800-482-4822 to order a replacement display board (P/N TC10012952 for LVP Display Board, P/N TC10003525 for Syringe/PCA Displa |
| Quantity | 89497 |
| Distribution | Worldwide distribution - U.S. Nationwide distribution including in the states of CO, FL, SC, NY, CA, NC, TX, IL, PA, AZ, LA, IN, MA, MO, AR, WI, HI, OR, OH, NJ, TN, NE, WA, KS, MI, MD, WV, KY, UT, MN, IA, GA, AL, DE, AK, VA, OK, NM, CT, SD, NV, ID, MT, ME, VT, MS, WY, NH, MP, DC, ND, GU, PR, AS, RI |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K133532 | ALARIS SYSTEM WITH GUARDRAILS SUITE MX Carefusion 303, Inc., Aug 21, 2014 |