ClearanceBrief
Recall Z-2822-2020 ยท FRN Pump, Infusion

CareFusion 303, Inc.

Alaris Pump Module Model 8100 - Product Usage: is intended to provide trained healthcare caregivers a way to automate the programming of infusion parameters, thereby decreasing the amount of manual steps necessary to enter infusion data.

DateAug 19, 2020 (initiated Jun 30, 2020)
StatusOpen, Classified
ReasonLED display, which provides infusion or patient monitoring values, on the module may have segments that appear dim.
Root causeDevice Design
ActionOn 06/30/2020, Urgent Medical Device Recall Notifications will be mailed to Directors of Biomedical Engineering, Nursing, Risk Management, Environmental Services. Actions for Clinical User: 1. If the end user observes a dim segment on the module, replace the module when feasible. There is no impact to the infusion. 2. Complete and return the Customer Response Card. Actions by Biomedical Engineering: If the module has a dim segment, please contactrecalling firm at 1-800-482-4822 to order a replacement display board (P/N TC10012952 for LVP Display Board, P/N TC10003525 for Syringe/PCA Displa
Quantity1442097
DistributionWorldwide distribution - U.S. Nationwide distribution including in the states of CO, FL, SC, NY, CA, NC, TX, IL, PA, AZ, LA, IN, MA, MO, AR, WI, HI, OR, OH, NJ, TN, NE, WA, KS, MI, MD, WV, KY, UT, MN, IA, GA, AL, DE, AK, VA, OK, NM, CT, SD, NV, ID, MT, ME, VT, MS, WY, NH, MP, DC, ND, GU, PR, AS, RI
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133532ALARIS SYSTEM WITH GUARDRAILS SUITE MX
Carefusion 303, Inc., Aug 21, 2014