Recall Z-2740-2020 ยท FRN Pump, Infusion

CareFusion 303, Inc.

Alaris System PC Unit Model 8000 modular infusion pump and monitoring system

DateAug 6, 2020 (initiated Jun 30, 2020)
StatusTerminated
ReasonIf one or more screws or washers are loose or missing causing the battery not to be properly secured to the PC Unit, that is running on battery power, the system may experience a power loss with a prolonged, non-silenceable alarm. Power loss may result in an interruption of patient therapy or monitoring.
Root causeNonconforming Material/Component
ActionOn 06/30/2020, Urgent Medical Device Recall Notifications were mailed to Directors of Biomedical Engineering, Nursing, Risk Management, Environmental Services. Actions for Clinical Users: This situation only occurs when the device is running on battery power. 1. If this issue occurs, immediately plug the PC unit into AC power and restart the infusions. 2. Obtain a replacement PC unit as soon as possible. 3. Ensure back-up devices are readily available when infusing critical medications. Please also ensure that you have back-up monitoring devices (e.g., EtCO2, SpO2) that are critically importa
Quantity37141
DistributionU.S.: CO, FL, SC, NY, CA, NC, TX, IL, PA, AZ, LA, IN, MA, MO, AR, WI, HI, OR, OH, KS, MN, MI, WA, UT, MD, TN, CO, NJ, NE, ND, WV, KY, IA, MS, AL, DE, AK, VA, OK, NM, GA, CT, SD, NV, ID, MT, ME, VT, WY, NH, MP, DC, GU, PR, AS, RI. O.U.S.: TR, QA, CA, AE, KW, MX, IL, SA, CH, NL, BE, TW, ZA, PH, CN,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133532ALARIS SYSTEM WITH GUARDRAILS SUITE MX
Carefusion 303, Inc., Aug 21, 2014