ClearanceBrief
Recall Z-2735-2020 ยท FRN Pump, Infusion

Smiths Medical ASD Inc.

Medfusion Syringe Pump Model 3500, Software Version V6.0. Infusion pump.

DateAug 6, 2020 (initiated Jun 26, 2020)
StatusOpen, Classified
ReasonInaccurate delivery can occur following an interrupted Bolus or Loading dose, if a specific sequence of events occurs.
Root causeSoftware design
ActionConsignees are being contacted via direct mailing and email. Consignees are asked to review and complete the Recall Response form indicating understanding of notification. Distributors are asked to forward the notification to parties to whom product has been distributed. Smiths Medical will send out 2 additional notifications by email. Phone calls will be made to non-responders.
Quantity17019 units
DistributionWorldwide distribution - United Arab Emirates, United States, Afghanistan, Albania, Australia, Belgium, Bermuda, Canada, China, Germany, Denmark, Great Britain, Hong Kong, Indonesia, Ireland, India, Korea, Kuwait, Lebanon, Mexico, Malaysia, New Zealand, Oman, Panama, Philippines, Pakistan, Poland, S
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K040899MEDEX 3000 SERIES MR SYRINGE INFUSION PUMP
Medex, Inc., Apr 30, 2004
K111386MEDFUSION MODEL 4000 SYRINGE INFUSION PUMP, PHARMGUARD TOOLBOX 2 MEDICATION SAFETY SOFTWARE AND SUPPORTED SYRINGES (PSS)
Smiths Medical MD, Inc., Aug 29, 2011