Recall Z-2880-2020 ยท DQK Computer, Diagnostic, Programmable

Biosense Webster (israel), Ltd.

CARTO 3 System (Software Version V7.1.80), CARTO VISITAG Module - Product Usage: FGThe intended use of the system is catheter-based cardiac electrophysiological (EP) procedures. The system provides information about the electrical activity of the heart and about catheter location during the procedure. The system can be used on patients who are eligible for a conventional electrophysiological proce

DateJun 22, 2020
StatusTerminated
ReasonSoftware defect may result in disapperance of tag sites during recalculation and lead to additional ablation sites. This may lead to prolongation of the procedure and, in extremely rare circumstances, cardiac perforation.
Root causeSoftware design
ActionOn 06/22/2020, the firm notified affected US customers via mail of a hospital letter, "URGENT Medical Device Correction", which indicates the following: This product is not being removed from the market and does not need to be returned. A software service pack will be provided by the firm. Your representative will notify you when it is available and will arrange for installation of the service pack. In the meantime, the firm has identified certain mitigations preventing the issue from occurring: If your system is equipped with the VISITAG SURPOINT Module License: - Always enable the VISITA
DistributionWorldwide distribution - US Nationwide distribution including in the states of AL, AK, AZ, AR, CA, CO, FL, GA, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WV, and WI *Three (3) U.S. government consignees. The cou
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K191660CARTO 3 EP Navigation System, Version 7.1 and Accessories
Biosense Webster, Inc., Jul 20, 2019