Recall Z-2697-2020 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

Philips North America, LLC

IntelliVue MX700 patient monitor, model no. 865241

DateJun 30, 2020
StatusTerminated
ReasonSome Intellivue MX700/MX800 Patient Monitors were delivered with defective power supplies. The defective power supplies might energize the metal parts of monitor if the grounding instructions were not followed, and may also cause the monitor to stop operating or prevent it from being switched on.
Root causeProcess control
ActionOn June 30, 2020, the firm sent Urgent Medical Device Correction letters to affected customers. Customers were instructed to always ensure that the monitor is grounded during operation with mains power. Affected products should be removed from use until the power supply of the monitor has been replaced. Philips Customer Service will contact customers when a replacement power supply is ready for installation. Philips expects all replacements to be complete within six months. If you need any further information or support concerning this issue, please contact your local Philips representative o
Quantity28
DistributionDomestic distribution to CA, HI, MA, NY, and UT. International distribution to Costa Rica, Ireland, Lebanon, and Poland.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K150310IntelliVue Patient Monitor
Philips Medizin Systeme Boeblingen GmbH, May 13, 2015