Recall Z-2699-2020 ยท JOJ Indicator, Physical/Chemical Sterilization Process

Resource Optimization & Innovation LLC

regard Migrating Indicator, STEAM Migrating Indicator Strips, REORDER # 335576REG - Product Usage: The integrating indicator is designed to chemically react over time with the critical parameters of steam sterilization cycle within a specified tolerance.

DateJun 17, 2020
StatusTerminated
ReasonThe firm received reports from customer that the Regard Migrating Indicator shown malfunctions and does not go into the "accept" window at the time of use.
Root causeUnder Investigation by firm
ActionThe firm initiated the recall by email on 06/17/2020. The letter requested the customer quarantine any product and notify the firm. ROi, LLC will facility the return of the product. The letter also directed the customer notify all potential users how may have received the recalled item.
Quantity66 cases of 12,000 units per case
DistributionUS Nationwide distribution including in the states of MO, NC, AR, OK.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K982188STERI-PAK BOWIE DICK TEST PACK
Steritec Products Mfg Co, Inc., Nov 20, 1998