ClearanceBrief
Recall Z-2695-2020 ยท GXI Probe, Radiofrequency Lesion

Medtronic Sofamor Danek USA Inc

Accurian RF Ablation,Cannula pk, Accurian, 100mm, 22G, 10mm, SS Product Number: AC0019 GTIN / UPN: 20763000153667 - Product Usage: (sterile, single-use) comprises an electrically insulated shaft, an active electrode for RF energy delivery, a hub with a female slip luer fitting, and a stylet with a color-coded cap to indicate gauge size. The ACCURIAN RF Cannula included in the ACCURIAN RF Ablation

DateJun 8, 2020
StatusTerminated
ReasonInsertion difficulty of the probes. The probe was catching on a lip inside the hub of the cannula.
Root causeUnder Investigation by firm
ActionMedtronic initiated both a voluntary correction notice pertaining to the labeling for the affected 20G and 22Gs and removal notice pertaining to the affected 18Gs on 10 June 2020. Notice of removal for ACCURIAN RF Cannula (18G, sterile, single-use) state the following action to be taken: 1) Locate and remove the impacted product from normal storage locations. Do not use this product. 2) Complete the enclosed Customer Confirmation Form and return via email to rs.sdmriskmgt@medtronic.com. Please return this form to Medtronic even if you have no affected inventory. 3) As applicable, contact you
Quantity25 pk
DistributionWorldwide distribution - US Nationwide distribution and the countries of Belgium and Australia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K181864Polaris RF Ablation System
Baylis Medical Company, Inc., Jan 2, 2019