Recall Z-2540-2020 ยท GZL Electrode, Depth

FHC, Inc.

FHC Guideline 4000 5.0 GL5 Main Processing Unit - Product Usage: intended to record and stimulate electrophysiological activity, as well as aid in the accurate placement of electrodes and other instruments. Catalogue Number: C0215 UDI: 00873263006740

DateJun 17, 2020
StatusTerminated
ReasonAn electrical sub-circuit was not properly connected. Its purpose is to eliminate the possibility of interference with proper performance of the Guideline 5 Main Processing Unit, if an ESD event (greater than 6kV) were to occur.
Root causeComponent change control
ActionFHC issued Urgent Medical Device Recall letter dated 6/17/20 via USPS certified return receipt,stating reason for recall, health risk and action to take: FHC, Inc. field representatives will be contacting you to return the effected product and issue a replacement. In the interim, we advise you to stop use of this product. Should you have any further questions regarding this matter you may contact me, by email at kmoeykens@fh-co.com or by telephone at 207-666-5425.
Quantity6 units
DistributionUS Nationwide distribution including in the states of CA, NC, TN, TX, WA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K183123microTargeting Guideline 4000 5.0 System
FHC, Inc., Dec 20, 2018