Recall Z-2475-2020 ยท ITQ Joint, Knee, External Brace

Mueller Sports Medicine, Inc

Mueller Sports Medicine Hybrid Wraparound Knee Support, Qty: 1 unit per box

DateJun 18, 2020
StatusTerminated
ReasonNatural rubber is present in Hybrid Wraparound Knee Support labeled as "Not made with Natural Rubber latex".
Root causeProcess change control
ActionMueller Sports sent notification letters to both customers and distributors dated, 6/12/2020, and titled, "URGENT MEDICAL DEVICE RECALL". The letter indicate that Natural Rubber was present in the Hybrid Wraparound Knee Support and the product is labeled with the statement: Not made with Natural Rubber Latex. The firm initiated a voluntary recall and is notifying customers of the issue. Customers should immediately discontinue use of any product identified in the letter. If customers have affected product, they need to complete the enclosed Recall Acknowledgment Form and fax to Mueller Spo
Quantity135,354 units
DistributionU.S. distribution only: AR and OH.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.