Recall Z-2983-2020 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Beckman Coulter Biomedical GmbH

DxA 5000; Catalog Number:B50516; UDI:15099590707118 - Product Usage: The DxA 5000 is a high-speed, modular, automated sample handling system that performs pre-analytical and post-analytical sample processing and storage. The automation system also sorts, routes, and presents sample tubes to analyzers for analysis. The DxA 5000 also consolidates a variety of analytical instruments, such as an Immun

DateJun 17, 2020
StatusTerminated
ReasonThe DxA 5000 instructions for use may be confusing regarding the re-use of sample IDs. Beckman Coulter has received complaints where the re-use of sample barcodes were not being detected by the system .
Root causeDevice Design
ActionThe Urgent Medical Device Recall letters were sent via first class mail on 06/19/20. The actions to be taken by consignees are as follows: Avoid Duplicate IDs; DO NOT Re-use Sample IDs As indicated in the DxA 5000 instructions for use, ensure that only unique sample barcode IDs are used. Be aware of the difference between duplicate sample IDs and the re-use of the same sample ID, as the instrument appropriately detects duplicates, but not re-used sample IDs. *Duplicate sample IDs are defined as samples with the same ID concurrently being run on the instrument. *Re-used sample IDs are defined a
Quantity32 units
DistributionWorldwide distribution - US Nationwide distribution including in the states of Illinois, Ohio, Pennsylvania, and West Virginia. The countries of Australia, Canada, China, Czech Republic, France, Israel, Italy, Germany, Switzerland, Spain, Netherlands, and United Kingdom.