Recall Z-2776-2020 ยท PEN Gram-Negative Bacteria And Associated Resistance Markers

Luminex Corporation

Verigene Processor SP with assay Luminex, VERIGENE Gram-Negative Blood Culture Nucleic Acid Test, BC-GN, IVD - Product Usage: utilizes a single-use disposable test consumables and a self-contained Verigene Test Cartridge for each sample tested.

DateJun 15, 2020
StatusTerminated
ReasonAn instrument failure mode that may result in a Blood Culture Gram Positive (BC-GP) or Gram Negative (BC-GN) false negative call.
Root causeDevice Design
ActionOn 06/10/2020, the firm sent a letter to customers titled, "MEDICAL DEVICE FIELD ACTION", regarding an instrument failure mode that may result in a Blood Culture Gram Positive (BC-GP) or Gram Negative (BC-GN) false negative call. The firm has developed a software solution that identifies and alerts users to the issue. The firm will contact customers to verify their address to ship the software material and provide installation information. Once the update has been completed, customers need to email a screen shot of the updated software, as well as a completed Notice of Destruction to Global S
QuantityTotal = 3,016 systems (2240 Verigene Processor SP and 776 Verigene Processor SP Refurbished)
DistributionWorldwide distribution - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CN, CT, DE, DC, FL,GA, ID, IN, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NB, NV, NE, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN TX, UT, VT, VA, WA, WV, WY, and WI. The countri
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K132843VERIGENE GRAM NEGATIVE BLOOD CULTURE NUCLEIC ACID TEST (BC-GN)
Nanosphere, Inc., Jan 8, 2014