ClearanceBrief
Recall Z-2668-2020 ยท PTI Non-Coring (Huber) Needle

Becton Dickinson & Company

SafeStep Huber Needle Set with GuardIVa Antimicrobial Hemostatic Dressing Port Access Kit; Catalog Numbers PA-0031G (UDI 00801741085666), PA-0031YNG (UDI 00801741085673), PA-0032G (UDI 00801741087639), PA-0033G (UDI 00801741096433)

DateJul 20, 2020 (initiated Jun 19, 2020)
StatusTerminated
ReasonPotential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that have developed breaks or cracks can cause leakage resulting in exposure to chemicals such a chemotherapeutics or biohazard material (e.g., blood). A crack or break could result in potential blood loss, catheter occlusion, air embolism, under or interrupted infusion, or site contamination which could lead to infection.
Root causeProcess control
ActionURGENT MEDICAL DEVICE RECALL notification letters dated 6/19/20 were sent to customers. PLEASE TAKE THE FOLLOWING ACTIONS: 1. Immediately review your inventory for the specific catalog and lot numbers referenced in Attachment A. Discontinue use of those devices and destroy all product subject to this recall following your institutions process for destruction. MDS-20-1979 Page 2 of 28 BD Medical 1 Becton Drive Franklin Lakes, NJ 07417 USA bd.com 2. Share this notification with all users of the product within your facility network to ensure they are also aware of this recall. 3. Complete the a
Quantity5,296,132 total
DistributionWorldwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K153440PowerLoc MAX Power Injectable Infusion Set, SafeStep Huber Needle Set
C.R. Bard, Inc., Aug 18, 2016