ClearanceBrief
Recall Z-2430-2020 ยท NDW Assay, Porphyrin, Spectrophotometry, Lithium

Siemens Healthcare Diagnostics, Inc.

ADVIA Chemistry Calibrator

DateJun 5, 2020
StatusTerminated
ReasonCalibration errors have been observed with the Total and Direct Bilirubin assays when using Chemistry Calibrator lot 534177 for ADVIA Chemistry and lot 534179 for Atellica CH.
Root causeNonconforming Material/Component
ActionSiemens Healthcare Diagnostics sent a letter notification sent to customers dated June 10, 2020. As one of the short-term alternatives for customers, where regulations permit, Siemens has evaluated the use of the lyophilized RANDOX Calibration Serum Level 3 Lot 1024UE. (CAL 3) for generating DBIL_2/DBil_2 and TBIL_2/TBil_2 results until a new lot of unaffected Siemens Chemistry Calibrator is available.
Quantity1872 units
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K050374ADVIA IMS LITHIUM ASSAY
Bayer Healthcare, LLC, Jun 15, 2005