ClearanceBrief
Recall Z-2672-2020 · NKB Thoracolumbosacral Pedicle Screw System

U&I CORP.

ANAX 5.5 Poly-Reduction Screw 6.0X50MM, REF: SFNA6050R, part of ANAX 5.5 SPINAL SYSTEM

DateMay 26, 2020
StatusTerminated
Reason6.0x50mm poly-reduction screws were laser etched 6.0x45
Root causeProcess control
ActionOn 05/26/20, a Request to Return Products Which Have Laser Marking Issue notice was sent to the firm customer. The customer was first asked to distinguish and isolate affected product and then to return affected product. This customer/distributor sent Notification for Sales Suspension and Product Recall notices to their customers. Customers were asked to return affected product and to fill out and return the Notification for Sales Suspension and Product Recall notices to acknowledge receipt.
Quantity6
DistributionU.S.: AZ
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K173524ANAX 5.5™ Spinal System
U&I Corporation, Dec 13, 2017