Recall Z-2551-2020 ยท JJH Clinical Sample Concentrator

Roche Molecular Systems, Inc.

MagNA Pure 96 Instrument, clinical sample concentrator, Material number 06541089001 - Product Usage: This protocol is intended for pipetting samples into a MagNA Pure 96 Processing Cartridge (purification rack position 1) before a purification run.

DateJun 3, 2020
StatusTerminated
ReasonWhen using Sample Transfer protocol version 3.0, the drop catcher is not activated on the MagNA Pure 96 instrument, which introduces a potential risk for cross-contamination leading to erroneous results.
Root causeSoftware change control
ActionOn June 5, 2020, the firm distributed Urgent Medical Device Correction letters to consignee accounts. The UMDC instructs the customers to determine whether or not their laboratory is using the affected Sample Transfer protocol version 3.0 with their MagNA Pure 96 instrument. If the affected protocol is found on the instruments Control Unit, it needs to be immediately deleted by following the instructions provided in the UMDC. A newly updated Sample Transfer protocol version 4.0 is available. Roche Affiliate organizations will provide instructions for obtaining the new version of the protocol.
Quantity917
DistributionWorldwide distribution - US Nationwide distribution.