ClearanceBrief
Recall Z-2526-2020 ยท LXH Orthopedic Manual Surgical Instrument

TeDan Surgical Innovations LLC

PHANTOM CS DISTRACTION SCREW, 12 MM, STERILE, 5/BOX - Product Usage: intended to be used for distraction in the anterior approach to the cervical spine.

DateMay 15, 2020
StatusTerminated
ReasonLack of pouch seal. Product is labelled as sterile.
Root causePackaging change control
ActionOn June 4, 2020, the firm sent an "URGENT: MEDICAL DEVICE FIELD SAFETY NOTICE" along a "Field Safety Notice Acknowledgement and Receipt Form Response Required" to customers via email informing them the Recaling Firm has received two (2) complaints associated with the packaging issue where there was no seal on the screw pouch opposite the chevron manufacturing pouch seal. The Recalling Firm is instructing customer to: -immediate visual inspection of the pouch be conducted (Letter A shown in the red outline opposite the chevron seal [Letter B] in Figure 1) and observe for an open seal (example
Quantity1965 screws (packaged into 393 boxes) - [5 screws per box]
DistributionWorldwide distribution - U.S. Nationwide distribution including in the states of: AZ, CA, FL, GA, HI, IL, MA, MI, NC, NH, NJ, NY, NV, PA, PR, TN, TX, and WA . The countries of Australia, Canada, France, Philippines, Taiwan, U.A.E., and United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.