ClearanceBrief
Recall Z-2506-2020 ยท NVZ Pulse Generator, Permanent, Implantable

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)

Patient Connector, Model Number 24967

DateJul 10, 2020 (initiated Jun 4, 2020)
StatusOpen, Classified
ReasonThere is a rare communication sequence during the first device interrogation with a SmartSync Device Manager that may result in the temporary suspension of some device features (i.e., battery measurements, Capture Management", Atrial Lead Position Check", EffectivCRT" algorithms, and AdaptivCRT"). This rare interaction results in temporary suspension of automatic threshold testing and output adjustments, and suspension of auto-optimization of CRT therapy. The issue is unlikely to result in clinical impact to the patient, and features are restored upon next programmer device interrogation or p
Root causeSoftware design
ActionThe firm initiated their medical device correction on 06/04/2020 by letter. The letter explained the problem and informed the consignee of the available software update (version 3.2.01) along with download instructions.
Quantity1025 devices
Distributionworldwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.