Recall Z-2523-2020 ยท JAA System, X-Ray, Fluoroscopic, Image-Intensified

Philips North America, LLC

EasyDiagnost Eleva DRF, model no. 706050 - Product Usage: X-ray system, diagnostic, general-purpose.

DateJul 10, 2020 (initiated Jun 2, 2020)
StatusTerminated
ReasonThermo switches in the main power supply for the system may be incorrectly installed, resulting in a malfunction.
Root causeProcess control
ActionThe firm distributed Urgent Field Safety Notice letters to affected customers on 6/2/2020. Customers were informed that the thermos switches may be incorrectly installed which results in a malfunction. If the transformer overheats due to a failure and the thermos switch is not activated, the device may begin to generate smoke. Customers were advised by the firm that they can use the system according to the IFU (Instruction for Use). The user should pay attention to whether it smells of smoke in the room. If this is the case, immediately shut down the system according to IFU or (if present) us
Quantity52
DistributionWorldwide distribution - US Nationwide distribution including in the states of AR, AZ, CA, CT, FL, GA, HI, IL, LA, MA, MD, MI, MS, MT, NC, NY, TN, TX and the country of Panama.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K031535PHILIPS EASYDIAGNOST ELEVA
Philips Medical Systems North America Co., Jun 17, 2003