Recall Z-2537-2020 ยท DRG Transmitters And Receivers, Physiological Signal, Radiofrequency

Vitalconnect Inc.

VistaSolution, physiological signal transmitter. A software graphical user interface for use by healthcare professionals to display physiological data collected by the wireless remote monitoring system in healthcare settings. This is a secondary, adjunct patient monitor and is not intended to replace existing standard-of-care patient monitoring.

DateJul 10, 2020 (initiated May 21, 2020)
StatusTerminated
ReasonA healthcare provider (HCP) noticed a patient had a lowered historic SpO2 reading that the software should have triggered a notification to the care team, but no notification was found. A software defect was encountered causing the lack of notification. The defect could result in exposing a patient to potential hypoxia for a short window of time. The firm is correcting the software to bring the performance back into compliance with specifications.
Root causeSoftware Design Change
ActionOn May 21, 2020, the firm sent a "Urgent: Medical Device Recall" notification to all affected customers via E-mail and mail. The letter stated, that the third party Sp02 continuous device notifications may not indicate an alert as intended. Therefore, the health care provider may not be notified to check the patient. Potential Hazards: A patient may have hypoxia and the designated health care provider may not be notified. The firm will be turning off the ability to set custom notifications containing continuous SP02 until the issue is resolved. The firm tells customers to monitor the patient
Quantity7
DistributionU.S. Nationwide distribution including in the states of NJ, NY, VA, OUS: None
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K193343Vista Solution Monitoring Kit, Vista Solution 2 10-day Evaluation Kit, VistaTablet 2 (Boxed)
Vitalconnect, Inc., Apr 20, 2020