Recall Z-2542-2020 ยท OMP Negative Pressure Wound Therapy Powered Suction Pump

KCI USA, INC.

ACTIV.A.C Therapy System; Model Numbers; US: 340000 ACTIV.A.C., NORTH AMERICAN UDI (01)00878237008188 - Product Usage: is an integrated wound management system for use in the acute, extended, and home care settings. When used on open wounds, it is intended to create an environment that promotes wound healing by secondary or tertiary (delayed primary) intention by preparing the wound bed for closur

DateJun 8, 2020
StatusTerminated
ReasonThe device may power off without notification to the user resulting in a stoppage of negative pressure wound therapy.
Root causeUnder Investigation by firm
ActionFor Customer Owned units, KCI plans to notify customers via customer letters delivered by common carriers or via email communication that the ACTIV.A.C." Therapy Unit may shut down without notification to the user. The customer letter notification or email communication will inform customers to coordinate the product software change. The recalling firm sent an URGENT - MEDICAL DEVICE RECALL letter to the affected customers on 06/08/2020. The completed Customer Response Form is to be returned to kci3mfieldactionresponse@mmm.com
Quantity5,823 units
DistributionWorldwide distribution - US Nationwide distribution including in the states of AZ, FL, IL, OH, MD, ND, NV, OK, SC, TN, TX, UT, VA, WA and Puerto Rico. The country of Australia, Austria, Canada, China, France, Germany, Japan, Hong Kong, India, Indonesia, Malaysia, New Zealand, Philippines, South Kor
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K183543ACTIV.A.C. Therapy Unit
Kci USA, Inc., Mar 20, 2019