Recall Z-2500-2020 ยท HWA Impactor

XTANT Medical

Irix-C Forceps Inserter; Product Code/UDI: T066-0059/ M697T06600591

DateMay 12, 2020
StatusTerminated
ReasonThe strike plate fractured off from the forceps inserter devices with minimal impaction or force due to unsatisfactory weld.
Root causeProcess design
ActionOn May 12, 2020 a customer letter (in an envelope marked URGENT) was Federal Expressed priority overnight-signature required to the three customers who received the affected devices. On May 19, 2020, an additional customer letter (in an envelope marked URGENT) was Federal Expressed priority overnight-signature required to one of the three affected customers.
Quantity4 units
DistributionAZ, TX and WV.