Recall Z-2474-2020 ยท MVR Device, Anastomotic, Microvascular

Baxter Healthcare Corporation

Flow Coupler Monitor, GEM, Global Excellence Microsurgery, Rx Only

DateJun 2, 2020
StatusOpen, Classified
ReasonA firmware issue may cause the GEM1020M-2 Flow Coupler Monitor to intermittently shut down when a WiFi connection is not established.
Root causeDevice Design
ActionSynovis Micro Companies Alliance (MCA), Inc., a subsidiary of Baxter International Inc., sent a notification to healthcare providers dated, June 2, 2020 titled, "Urgent Medical Device Recall". The notification was sent to the user level for the GEM1020M-2 Flow Coupler Monitor due to a firmware issue which may cause the monitor to intermittently shut down when a WiFi connection is not established. If a shutdown occurs, the monitor can be manually powered back on and will resume normal operation. Synovis is requesting the return of the monitors to perform a firmware upgrade to resolve the issue
Quantity60 units
DistributionU.S. distribution only: VA, GA, MD, PA, DC, WI, LA, MN, FL, IN, MI, AZ, TX, MS, MA, NY, SC, WI, WA, MI, CA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K191252Gem FlowCoupler System
Baxter/ Synovis Micro Companies Alliance, Inc., Jul 9, 2019