ClearanceBrief
Recall Z-2259-2020 ยท KPS System, Tomography, Computed, Emission

GE Healthcare, LLC

System, Tomography Computed Emmission

DateJun 3, 2020 (initiated Nov 30, 2018)
StatusOpen, Classified
ReasonGE Healthcare has identified that a small number of Nuclear Medicine systems are operating without currently updated software. In a recent event, a patient s finger was injured during the Unload process while the automatic Unload feature was being used; the software on the system was not current.
Root causeOther
ActionGE Healthcare identified that a small number of Nuclear Medicine systems are operating without currently updated software.
Quantity526 total units
DistributionDomestic Distribution: AL, AR, AZ, CA, CO, CT, DC, DE, FL,GA, IA, IL, IN, KY, LA,MA, MD, ME, MI, MN, MO, MS, NC, NE, NM, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT,VA, WI and WV. Foreign Distribution: Argentina, Australia, Belgium, Brazil, Canada, Colombia, Denmark, France, Germany, Ghana, Indi
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K080124VENTRI 1.1
Ge Medical Systems F.I. Haifa, Jan 31, 2008