ClearanceBrief
Recall Z-2424-2020 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

GE Healthcare, LLC

CARESCAPE Patient Data Module (PDM) used with the following monitors: CARESCAPE B450/B650/B850, SOLAR 8000M/I and Transport Pro, Model Numbers: 2042084-001 2094504-001 2042084-001-046050 2042084-001-208621 2042084-001-045915 2042084-001-215164 2042084-001-078474 2042084-001-053034 2042084-001-215165 2042084-001-079462 2042084-001-047389 2042084-001-112198 2042084-001-041673 2042084-001-227238 2042

DateJun 24, 2020 (initiated May 18, 2020)
StatusOpen, Classified
ReasonThe CARESCAPE Patient Data Module (PDM) is used with the following monitors: CARESCAPE B450/B650/B850, SOLAR 8000M/I and Transport Pro. If Pace detection is turned ON, on the monitor, and if an Automated External Defibrillator (AED) is used to perform defibrillation, the low frequency signals deployed by the AED, cause the PDM to issue a false flat line ECG waveform and alarm Asystole to be displayed on the monitor.
Root causeSoftware Design Change
ActionThe firm notified its consignees by letter on 05/18/2020. The letter explained the issue, how it occurs, cautions and a workaround. The notices were disseminated to the following entities: Director of Biomedical/Clinical Engineering, Chief of Nursing, Health Care Administrator/Risk Manager.
Quantity89,109 units
DistributionWorldwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K071073SOLAR 8000 AND TRANSPORT PRO WITH PATIENT DATA MODULE
Ge Medical Systems Information Technologies, May 11, 2007