Recall Z-2406-2020 ยท LHN System, Radiation Therapy, Charged-Particle, Medical

Mevion Medical Systems, Inc.

MEVION S250i, MEVION S250; Proton Radiation Treatment System

DateMay 15, 2020
StatusTerminated
ReasonCouch Correction moves sent after using a 3D CT scan are partially lost if the previous correction sent or recorded had some of the Couch Correction checkboxes turned off
Root causeSoftware design
ActionMevion issued letter (pCAR0012) on May 15, 2020 via mail or FedEx, with return receipt confirmation. Letter states reason for recall, health risk and action to take: Until a software fix is applied for this problem, the user is cautioned to carefully pay attention to the 3 indicators of Couch corrections. If they disagree with the intended moves the user may simply repeat the alignment step and the problem should clear. Required User Action When using multimodal imaging the user should always check the indicators for Couch Corrections before they are applied. This includes the Verity Couch Cor
Quantity8 units
DistributionDistribution to US in MO, NJ, OH, OK, FL, and DC, and the Netherlands
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K120676S-250 PROTON BEAM RADIATION THERAPY SYSTEM
Mevion Medical Systems, Jun 4, 2012
K172848MEVION S250i Proton Beam Radiation Therapy Device
Mevion Medical Systems, Dec 27, 2017