ClearanceBrief
Recall Z-2344-2020 ยท MUJ System, Planning, Radiation Therapy Treatment

RAYSEARCH LABORATORIES AB

RayStation 8A, UDI # 07350002010112 - Product Usage: For radiation therapy and medical oncology which can also be used to administer treatments.

DateMay 18, 2020
StatusTerminated
ReasonSoftware error was identified in the Pencil Beam Scanning (PBS) and Line Scanning (LS) dose calculation algorithms that could result in a local underestimation of expected dose.
Root causeSoftware design
ActionOn May 18, 2020, RaySearch Laboratories emailed an Urgent Field Safety Notice, Medical Device Correction #60564 letter to their customers. The letter listed the following actions to be taken by the users: always use the MC dose engine for final dose calculation, in robustness analysis, and dose tracking if possible (the MC dose engine is available in versions RayStation 6.0 and later); if the pencil beam dose engine is used for final dose calculation, ensure that all spots intersect the External ROI as seen in the Beams Eye View (BEV) of the Plan optimization module; if they do not, expand the
DistributionUS Nationwide distribution including in the states of Florida, Georgia, Utah, Virginia, California, Maryland, Arizona, Texas, Washington DC, Missouri, Massachusetts, Illinois, Oklahoma, New Jersey, Tennessee, and Louisiana.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K190387RayStation
RaySearch Laboratories AB (PUBL), Jun 19, 2019