Recall Z-2308-2020 ยท HRS Plate, Fixation, Bone

Biomet

A.L.P.S. T15 LOW PROFILE CORTICAL SCREW 3.5mm x 28mm, Item Number: 110017628 - Product Usage: Intended for temporary internal fixation and stabilization of osteotomies and fractures during the normal healing process.

DateMay 14, 2020
StatusTerminated
ReasonDistributed to the field without having completed design history files or design transfer activities
Root causeDevice Design
ActionZimmmer Biomet issued Urgent Medical Device Recall letter , dated May 14, 2020 to Risk Manger, Distributors,Surgeons, stating reason for recall, health risk and action to take: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbi
Quantity15 OUS
DistributionWorldwide distribution - US Nationwide distribution including in the states of NY, GA, FL, MO and the countries of Canada, AUSTRALIA, CHILE, CHINA, JAPAN, MALAYSIA, NETHERLANDS.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K083843LOCKING ANATOMIC & COMPOSITE PLATING SYSTEM
DePuy Orthopaedics, Inc., Mar 20, 2009
K173767A. L. P. S. Clavicle Plating System
Biomet, Inc., Apr 6, 2018