ClearanceBrief
Recall Z-2252-2020 ยท GCI Laryngoscope, Endoscope

Covidien Llc

Multifire VersaTrack Auto Suture Hernia Stapler, 4.0mm, Part No. 174021 - Product Usage: has application in endoscopic general surgery procedures for fixation of mesh in the repair of hernial defects and in other surgical specialties for approximation of tissue(s).

DateMay 12, 2020
StatusOpen, Classified
ReasonThe product is recalled due to a manufacturing assembly error that may cause the staples to fail to properly advance within the staple cartridge when the instrument handle is compressed.
Root causeProcess control
ActionAll consignees were notified by letter titled "Urgent Medical Device Recall" delivered via Federal Express and/or certified mail beginning May 12, 2020. The letter informs customers of the recall of specific item codes and production lots of Multifire hernia staplers due to a manufacturing assembly error. Use of a device with this condition may result in the failure of staples to properly advance within the staple cartridge when the instrument handle is compressed. Customers are instructed to take the following actions: " Immediately quarantine and discontinue use of the affected item code an
Quantity12,610 total
DistributionWorldwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K912097AUTO SUTURE ENDOSCOPIC FASCIA STAPLER
United States Surgical, A Division of Tyco Healthc, Sep 19, 1991