Recall Z-2355-2020 ยท OVD Intervertebral Fusion Device With Integrated Fixation, Lumbar

Acuity Surgical Devices, LLC

A LINK Z INTERVERTEBRAL BODY FUSION DEVICE Product Description: A Link Z Ti Unitary Open Interbody 32x24x12 7 A Link Z Ti Unitary Open Interbody 32x24x12 12 A Link Z Ti Unitary Open Interbody 32x24x16 7 A Link Z Ti Unitary Open Interbody 32x24x16 12 Product Usage: is a spinal device that is implanted in the intervertebral body space via an anterior approach to improve stability of the spine whil

DateJan 7, 2019
StatusOpen, Classified
ReasonDue to a manufacturing process issue, it allowed vertebral body bone screws to not seat properly within the spinal device/cage.
Root causeProcess control
ActionOn 12/14/2018, the firm notified its supplier via email requesting to visually review its inventory which determined that some screw holes were oval shaped. On 7/23/2019, all inventory was quarantined as part of the removal and was scrapped from inventory. Due to the low risk to patients and low probability, the firm ownly initially begin removal of product from internal locations (main warehouse) for the purpose of inspection and if necessary, repair or destruction. Removal Summary No additional customer or distributor communication took place during the removal.
QuantityA total of 27 lots producing 178 units
DistributionUS Nationwide distribution including in the states of TX, FL, NC, LA, KY, WI, SD, IL, CA, and KS. O.U.S.: None
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133827A-LINK Z
Osteo Vasive, LLC, Nov 17, 2014