Recall Z-2243-2020 ยท IYE Accelerator, Linear, Medical

Med Tec Inc

Type S Extension For use with Varian, CIVCO, Prescription Only - Product Usage: Device is intended for patient positioning.

DateJun 2, 2020 (initiated May 12, 2020)
StatusTerminated
ReasonPotential for disengagement of the Varian Style Type-S Extension during patient repositioning. The risk associated with the disengagement of the extension could be a patient fall if the patient is not supported by clinical staff.
Root causeComponent design/selection
ActionOn May 14, 2020, the firm sent a notification letter titled, "Urgent: Field Safety Notice Customer/Distributor Communication," to customers and distributors. The customer notification stated that a new warning will be added to the instruction manual at the next revision. The warning will state: To prevent extension from unintended disengagement, do not lift patient vertically to reposition while immobilized under a thermoplastic mask. Extensions that were further distributed, should be notified. An acknowledgement form should be returned to the firm by 05/29/2020. In addition, customers
Quantity171 units
DistributionWorldwide distribution - US Nationwide distribution including in the states of WI, TX, AZ, FL, NE, TN, MA, FL, GA, IN, OR, PA, IL, LA, NH, MS, MI, OH, MN, VA, IA, CA, NC, SC, AK, UT, WA, MO, MT, KS, and WV. The countries of Africa, Colombia, Saudi Arabia, Costa Rica, South Korea, Hungary, Spain, an
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K933227UNI-FRAME HEAD IMMOBILIZATION SYSTEM
Medtec, Inc., Jul 18, 1994