ClearanceBrief
Recall Z-2365-2020 ยท CIW Starch-Dye Bound Polymer, Amylase

Ortho-Clinical Diagnostics

VITROS Chemistry Product AMYL Slides Reagent, AMYLASE/5 PACK/90 SLDS, Product Code 8112724, UDI # 10758750004393 - Product Usage: For in vitro diagnostic use only to quantitatively measure amylase activity in serum, plasma, and urine using VITROS 250/350/5, 1 FS/4600/XT 3400 Chemistry Systems and VITROS 5600/XT 7600 Integrated Systems.

DateMay 11, 2020
StatusTerminated
ReasonMultiple coatings of VITROS Chemistry Products AMYL Slides are showing imprecision on control fluids when compared to the expected within lab SD as listed on the Performance Verifier (PV) Assay Sheets.
Root causeUnder Investigation by firm
ActionOn May 11, 2020, Ortho Clinical Diagnostics, Inc. sent an Urgent Medical Device Correction Notification to their customers alerting them of important information regarding the precision of VITROS AMYL Slides. The notice stated that the imprecision affects serum, urine and QC samples, however VITROS AMYL Slides performance continues to meet its intended clinical use. The VITROS Performance Verifiers I and II Control Assay Sheet estimated within-lab SD, Range of Means (ROM) and Delta table values were re-evaluated. The Delta tables will be updated in a future revision of the Performance Verifier
Quantity64,697
DistributionWorldwide distribution - US Nationwide including in the statesof Maine, Massachusetts, Vermont, Connecticut, New Jersey, New York, Pennsylvania, Washington DC, Maryland, Virginia, West Virginia, North Carolina, South Carolina, Georgia, Florida, Alabama, Tennessee, Missouri, Mississippi, Kentucky, Oh
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K812027KODAK EKTACHEM CLIN. CHEM. SLIDES (AMYL)
Eastman Kodak Company, Aug 3, 1981