ClearanceBrief
Recall Z-2558-2020 ยท FPA Set, Administration, Intravascular

CME America, LLC

BodyGuard Microset, REF: A120-160XSFK

DateJul 15, 2020 (initiated Jun 16, 2020)
StatusTerminated
ReasonInfusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy
Root causeDevice Design
ActionOn 06/16/20, Urgent Medical Device Recall Correction and Removal notices were emailed and mailed to customers and distributors. These notices provided additional information about the administration sets that were first identified in another recall notification issued on 04/27/20: www.bd.com/en-us/support/recall-notifications/recall-notification---cme-america-bodyguard-infusion-pump-system Customers ere asked to take the following immediate actions: 1) Share this letter with all users of the product within your facility to ensure awareness and understanding of this notice. 2) Identify any rema
Quantity273700
DistributionU.S: CA, IN, LA, MO, FL, IL, ME, MN, CO, VA, OK, NE, GA, AL, KY, IA, MS, ID, TX, SC, OH, SD, NV, MI, WY, NC, PA, TN, KS, NJ, OR, AR, NM, MA, NY, DC, RI, MD, WA, WI, AZ, NH, UT, HI, AK, MT, ND, VT, CT. OUS (Foreign): Canada, UK
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K042696BODYGUARD INFUSION PUMP SYSTEM
Caesarea Medical Electronics , Ltd., Mar 11, 2005