Recall Z-2302-2020 ยท OWB Interventional Fluoroscopic X-Ray System

Siemens Medical Solutions USA, Inc

Multi Display Manager in the Artis Systems (Image-Intensified Fluoroscopic X-Ray System) as follows: Artis Q floor Model: 10848280; Artis Q ceiling Model: 10848281; Artis Q biplane Model:10848282; Artis zee ceiling Model: 10094137; Artis zee floor Model: 10094135 Product Usage: Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging a

DateApr 30, 2020
StatusTerminated
ReasonArtis systems containing a Multi Display Manager of a specific lot, the power distributor may lead to an increased electrical contact resistance or the interruption of electrical contact, and result in planned procedures to be terminated and performed on an alternative x-ray system
Root causeNonconforming Material/Component
ActionSiemens Healthineers issued Urgent Medical Device Correction via AX036/19/S, dated 4/30/20, letter states reason for recall, health risk, and action to take: Siemens will exchange the affected power distributor of the Multi Display Manager. We advise you not to move the Multi Display Manager Container or to open the power distributor before the corrective action can be completed. A service organization will contact you to schedule an appointment to perform the correction via Update Instruction AX035/19/S. Please feel free to contact service organization at 1-800-888-7436 for an earlier app
Quantity15 US
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K181407Artis zee/zeego & Artis Q/Q.zen
Siemens Medical Solution USA, Inc., Aug 15, 2018