Recall Z-2236-2020 ยท FMZ Incubator, Neonatal

GE Healthcare, LLC

Giraffe Incubator with installed Servo Oxygen module. Infant incubator.

DateApr 28, 2020
StatusTerminated
ReasonPotential for certain Giraffe Incubators and Giraffe OmniBeds to deliver a different environmental oxygen level than what is displayed by the device, if a certain optional component referred to as the Servo Oxygen Module is installed.
Root causeDevice Design
ActionGE Healthcare notified customers on about 04/28/2020 via "URGENT MEDICAL DEVICE CORRECTION" letter. Customers were informed that certain Giraffe Incubators and Giraffe OmniBeds manufactured before 2012 can deliver a different environmental oxygen level than what is displayed by the device, if a certain optional component referred to as the Servo Oxygen Module is installed in the device. Instructions included that if the Giraffe Incubator or Giraffe OmniBed device was manufactured before 2012 AND has the oxygen sensor part number M-43GE installed, the Servo Oxygen feature will need to be di
Quantity5,883 units total
DistributionWorldwide Distribution. US nationwide including US Virgin Islands, Australia, Austria, Bahrain, Belgium, Bermuda, Bulgaria, Burundi, Canada, Chile, China, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Germany, Gree
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K010222OHMEDA MEDICAL GIRAFFE INCUBATOR
Ohmeda Medical, Mar 21, 2001
K020547OHMEDA MEDICAL GIRAFFE INCUBATOR
Ohmeda Medical, May 9, 2002
K101778GIRAFFE INCUBATOR
Ohmeda Medical, Jul 21, 2010
K993407OHMEDA MEDICAL OMNIBED
Ohmeda Medical, Jan 5, 2000