Recall Z-2285-2020 ยท JFY Enzymatic Method, Creatinine
Siemens Healthcare Diagnostics, Inc.
Dimension Vista Enzymatic Creatinine, Siemens Material Number 10700444, UDI Number 00842768036262 - Product Usage: use in the quantitative determination of creatinine in human serum, plasma, and urine for the diagnosis and treatment of renal disease.
| Date | Jun 5, 2020 (initiated May 4, 2020) |
| Status | Terminated |
| Reason | The presence of N-acetyl-p-benzoquinone imine (NAPQI), a metabolite of acetaminophen, may cause interference in certain chemistry assays when testing patient samples. |
| Root cause | Device Design |
| Action | On May 4, 2020, Siemens notified their customers that use the Fructosamine and Enzymatic Creatinine assays on the Atellica, ADVIA, and Dimension platforms of the potential for falsely depressed results in the presence of NAPQI. An Urgent Medical Device Correction Notice was sent to consignees in the US and an Urgent Field Safety Notice was sent to customers outside the US. The letter stated that the Limitations of the Procedure section of the Instructions for Use (IFU) will be updated with specific information related to the assay/platform. Customers were asked to complete and return the Effec |
| Quantity | 23,824 |
| Distribution | Worldwide distribution - US Nationwide distribution including in the states of North Carolina, California, Indiana, Florida, Ohio, Arizona, Massachusetts, New Jersey, New Hampshire, Illinois, Minnesota, Arkansas, New York, Delaware, Nevada, Kentucky, Washington DC, Montana, Missouri, Colorado, Maryl |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K090330 | DIMENSION VISTA ENZYMATIC CREATININE FLEX REAGENT CARTRIDGE, AND ENZYMATIC CREATININE CALIBRATOR, MODELS K1270, KC270 Siemens Healthcare Diagnostics, Inc., May 22, 2009 |