ClearanceBrief
Recall Z-2282-2020 ยท JFY Enzymatic Method, Creatinine

Siemens Healthcare Diagnostics, Inc.

ADVIA Chemistry Enzymatic Creatinine_2, Siemens Material Number 10335869, UDI Number 00630414526690 - Product Usage: use in the quantitative determination of creatinine in human serum, plasma, and urine for the diagnosis and treatment of renal disease.

DateJun 5, 2020 (initiated May 4, 2020)
StatusTerminated
ReasonThe presence of N-acetyl-p-benzoquinone imine (NAPQI), a metabolite of acetaminophen, may cause interference in certain chemistry assays when testing patient samples.
Root causeDevice Design
ActionOn May 4, 2020, Siemens notified their customers that use the Fructosamine and Enzymatic Creatinine assays on the Atellica, ADVIA, and Dimension platforms of the potential for falsely depressed results in the presence of NAPQI. An Urgent Medical Device Correction Notice was sent to consignees in the US and an Urgent Field Safety Notice was sent to customers outside the US. The letter stated that the Limitations of the Procedure section of the Instructions for Use (IFU) will be updated with specific information related to the assay/platform. Customers were asked to complete and return the Effec
Quantity1006
DistributionWorldwide distribution - US Nationwide distribution including in the states of North Carolina, California, Indiana, Florida, Ohio, Arizona, Massachusetts, New Jersey, New Hampshire, Illinois, Minnesota, Arkansas, New York, Delaware, Nevada, Kentucky, Washington DC, Montana, Missouri, Colorado, Maryl
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K070727ADVIA CHEMISTRY ENZYMATIC CREATININE_2
Siemens Medical Solutions Diagnostics, Aug 3, 2007