ClearanceBrief
Recall Z-2332-2020 ยท MBL Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Porous

Smith & Nephew, Inc.

smith&nephew R3 ZERO HOLE HA COATED HEMISPHERICAL STIKTITE COATED SHELL, manufactured in the following sizes: Size REF/product number 42MM 71332242 46MM 71332246 48MM 71332248 50MM 71332250 52MM 71332252 54MM 71332254 56MM 71332256 58MM 71332258 60MM 71332260 62MM 71332262 64MM 71332264 Orthopedic implant component.

DateApr 22, 2020
StatusOpen, Classified
ReasonA manufacturing error resulted in out of specification R3 Acetabular Shells
Root causeProcess control
ActionThe firm notified its consignees by letter (delivered by Fed Ex) and email beginning on 04/22/2020. Consignees are asked to inspect inventory, locate and quarantine affected devices. Affected product will be returned to Smith & Nephew. Questions or concerns regarding this recall may be directed to: FieldActions@smith-nephew.com.
Quantity1450 devices
DistributionWorldwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K070756SMITH AND NEPHEW REFLECTION 3 ACETABULAR SYSTEM
Smith & Nephew, Inc., Jun 6, 2007