ClearanceBrief
Recall Z-1994-2020 ยท KOD Catheter, Urological

Covidien Llc

Dover 400 ml 100% Silicone Silver-Hydrogel Coated Premium Urine Meter Foley Tray, 18 Fr/Ch (6.0 mm), 5 mL Catheter Pre-connected to 2000 mL Drainage Bag with Luer-Lock Sampling, Temperature Sensor and Adhesive Catheter Securement, Sterile, Product Code: P4P18XTSD - Product Usage: for Drainage of the Urinary bladder and simultaneous monitoring of temperature. Prescription use only.

DateApr 30, 2020
StatusTerminated
ReasonTemperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/- 0.2C. Failure to correctly measure patient body temperatures could lead to an incorrect diagnosis and insufficient treatment
Root causeNonconforming Material/Component
ActionCardinal Health issued letter Urgent Medical Device Recall EVENT-2020-02846 on 4/30/20 via overnight mail. Letter states reason for recall, health risk, and action to take: 1. CHECK all storage and usage locations to confirm whether you have any units of the affected product codes and lot numbers 2. SEGREGATE and QUARANTINE all on-hand product that is confirmed to be labeled per Exhibit A. 3. PLEASE RETURN the enclosed acknowledgment form via facsimile (847-689-9101 or 614-652-9648) or email (GMB-FieldCorrectiveAction@cardinalhealth.com) and indicate the product code, lot and quantity of produ
Quantity0
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K933400KENDALL CURITY 100% SILICONE FOLEY CATHETER W/TEMPERATURE SENSOR
Kendall Healthcare Products Co. Div.Of Tyco Health, Nov 15, 1993