Recall Z-2389-2020 ยท HRX Arthroscope
Medtronic Sofamor Danek USA Inc
Extremities, Large Kit Assy, US, 15/3 - Product Usage: is an injectable, moldable and biocompatible calcium phosphate bone void filler. The single-use Medtronic Bone Void Filler Kit contains the necessary components for mixing of the bone void filler.
| Date | Apr 24, 2020 |
| Status | Terminated |
| Reason | Product may not maintain its setting characteristics for the labeled shelf-life duration. Use may lead to longer time for product to set at the time of implantation which may increase the required surgical and anesthesia time for the patient. If the device seeps or creeps from its intended position prior to setting and creates incomplete void or gap filling, possible longer-term health consequences could include incomplete bone formation or lack of bone formation, delayed union, or nonunion. |
| Root cause | Under Investigation by firm |
| Action | Medtronic notified customers on about 04/24/2020 via "URGENT: Voluntary Market Removal (Recall)" letter. Customers were informed that Medtronic was notified by Skeletal Kinetics LLC that product that may not maintain its setting characteristics for the labeled shelf-life duration of the product, which may lead to a longer period of time required for the product to set at the time of implantation. Instructions included to not use the product and locate and removed the impacted product from inventory, complete the provided Customer Confirmation Form and return it via email to rs.sdmriskmgt@m |
| Quantity | 41 units |
| Distribution | Nationwide distribution to CA, FL, IL, IN, KY, MN, MO, MT, OH, PA, TX, VA, WA. |
| FDA record | Open on fda.gov |