Recall Z-2256-2020 ยท JAA System, X-Ray, Fluoroscopic, Image-Intensified

Philips North America, LLC

ProxiDiagnost N90, UDI 706100

DateApr 23, 2020
StatusTerminated
ReasonThe thermo switches in the three-phase transformer, which normally aid in powering down the System Power Distribution Unit, may be incorrectly installed and not working. If the transformer overheats due to a first failure and the thermo switch is not activated, the device may begin to generate smoke.
Root causeProcess control
ActionOn April 23, 2020, the firm distributed Urgent Field Safety Notice letters to affected customers. Customers were informed of the product issue. According to the notice, the customer can use the system according to the IfU (Instruction for Use). The user should pay attention to whether it smells of smoke in the room. If this is the case, immediately shut down the system according to IFU or (if present) use emergency off switch in room, remove patient, stop using the system for further examinations and contact customer service. Philips plans to perform an On-Site check of the SPDU thermo con
Quantity91
DistributionDomestic distribution nationwide. Foreign distribution to Chile and Germany.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K173433ProxiDiagnost N90
Philips Medical Systems Dmc GmbH, Feb 5, 2018